Operation overview
4.10 SUCTIONING
4.10.1 SUCTION SUPPORT
The Suction Support function makes it
possible to automatically inhibit the ventilator
from cycling during a tracheal suction
procedure without activating alarms.
Suction Support includes:
pre-oxygenation/preparation phase
disconnect phase
post-oxygenation phase
WARNINGS!
Suction Support is not intended to be
used together with closed-suction
systems.
The minimum PEEP level during suction
support is 3 cmH
adjust to the minimum level if the PEEP
level is below 3 cmH
disconnection of the patient.
Important: Alarms are turned off during
the disconnect phase for a maximum of 60
seconds. If the patient has not been
reconnected within 60 seconds, alarms are
activated.
56
|
O. The ventilator will
2
O in order to detect
2
Notes:
Suction Support is not available in NIV
mode or when the O
activated.
During the disconnect phase in Suction
Support, the nebulizer is temporarily
paused.
When only one gas is connected, an
elevated oxygen level cannot be set
during the preparation phase. In this
case, the post-oxygenation phase will be
skipped.
4.10.2 PRE-OXYGENATION/PREPARATION
PHASE
SVS-9000 XX
To enter the pre-oxygenation/preparation
phase:
1. Press the Quick access fixed key.
2. Press the Suction Support touchpad.
3. Set the desired pre-oxygenation value by
turning the Main Rotary Dial and press
Accept.
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
Breaths function is
2