3.3.2 STATUS MESSAGES
Message (message type)
Technical error no. 1 - 6, 29, 10001 (technical error)
Check battery status (Text Message)
Battery operation (Medium Priority Alarm)
Limited battery capacity (High Priority Alarm)
No battery capacity (High Priority Alarm)
Low battery voltage (High Priority Alarm)
3.3.3 AC POWER FAILURE
In the event of an AC power failure or
disconnection, the ventilator switches to
battery operation and activates medium
priority alarms, see Table above in Status
Messges for details.
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
Explanation
Power failure.
There is a problem with the battery modules. One
or more battery modules must be replaced.
AC power is off line due to a power failure or
disconnection.
Less than 10 minutes left of battery operation.
Less than 3 minutes left of battery operation.
Battery voltage too low. Cannot guarantee
continued ventilator operation.
|
Power supply
| 3 |
39