Operation overview
3. Start the Pre-use check by pressing Yes.
SVS-5073_XX
4. Follow the on-screen instructions.
5. After pressing the touchpad Pre-use
check the message Do you want to start
a pre-use check? is displayed - confirm
by pressing Yes.
44
|
4.2.2 INTERNAL LEAKAGE TEST
Connect the test tube between the inspiratory
outlet and the expiratory inlet.
Important: Use only the MAQUET test tube.
4.2.3 BATTERY SWITCH TEST
The Pre-use check tests the ventilator's ability
to switch between AC and battery power when
AC power is lost and restored:
When the on-screen instruction appears,
disconnect the ventilator from AC power.
When the on-screen instruction appears,
reconnect the ventilator to AC power.
Note: This test will not be performed if there
is less than 10 minutes of battery operation
available.
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
SVS-0082_XX