Start-up configuration
10.4 START MCARE REMOTE
SERVICE
Connect the network cable between the
SERVO-s Ventilator System (1) and the
docking station.
WARNING! Only use network equipment
that has been installed by a MAQUET
representative.
CAUTION: When using the MCare Remote
Service function, install the network cable
so that there is no risk of anyone tripping
over it.
Activate the MCare Remote Service function
by pressing the Start MCare Remote Service
touchpad (2) in the Biomed window.
2
The following message appears in Standby
(3) - MCare Remote Service is activated. Make
sure the network cable is connected and
Progress: Waiting for transfer...., followed by
Sending log files.
142
|
3
1
When the transfer is finished, the message
The file transfer is completed. MCare Remote
Service will be deactivated. Disconnect the
network cable.
Accept the dialog by pressing OK and the
MCare Remote Service is deactivated.
WARNING! Always disconnect the network
cable before starting ventilation.
Note: The SERVO-s Ventilator System is
prepared for the MCare Remote Service
functionality, although additional equipment
is needed to utilise this function. Please
contact your sales and service
representative for more details.