| 13 |
Appendix • User interface
13.3 QUICK ACCESS KEY
Suction Support
Loops
Waveform scales
Waveform
configuration
168
|
This function allows the user to disconnect the patient from the ventilator and turns
off the alarms (for apnea, minute volume, frequency alarm, check tubing, low PEEP)
during suction support.
Note: Suction support is not available when NIV or O
Loops are graphs that show two measured values: one measured value (x-axis)
against another measured value (y-axis). Loops are updated breath by breath.
Two loops are available:
volume-pressure
flow-volume
Waveform scaling
Pressure scaling
Flow scale
Volume scale
These three scales are set to automatic scaling, by default.
The sweep speed (mm/s) can be set to 10 or 20 mm/s (default).
Possibility to increase the space for viewing the waveform curves. This means that
more detailed information can be viewed.
breaths are activated.
2
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual