GETINGE GROUP Maquet SERVO-s VENTILATOR SYSTEM V7.0 User Manual page 12

Table of Contents

Advertisement

| 1 |
Introduction
|
1.2.4 GASES
CAUTION: The system is not intended to
be used with any anesthetic agent.
Important:
Supplied gases shall meet the
requirements for medical grade gases
according to applicable standards.
Maximum levels:
Air
H
O < 7 g/m
2
Oil < 0.5 mg/m
Chlorine: must not be detectable
Oxygen
H
O < 20 mg/m
2
Oil < 0.3 mg/m
Note: For devices with serial numbers
below 25000 that are updated to version
7.0, the maximum inlet gas pressure will be
reduced (see the Technical data chapter on
page 146).
1. If the compressed air is generated by a liquid ring compressor there is a potential risk of chlorine in the supplied
air.
12
3
3
3
3
1.2.5 AUXILIARY EQUIPMENT
CAUTIONS:
Accessories, supplies, and auxiliary
equipment used with the ventilator
should:
- be recommended by MAQUET
- meet EN/IEC 60601-1standards
- meet IEC standards as a whole system
If a scavenging system (i.e. gas
evacuation) is connected to the ventilator,
it must conform to ISO8835-3 guidelines
for subatmospheric pressure and
induced flow.
1
Measurements of parameter values that
have been processed by auxiliary
equipment:
- may be inaccurate if equipment not
authorized by MAQUET is used
- should be discounted if they conflict
with information on the ventilator
screen
- must not substitute for therapeutic or
diagnostic decisions
Note: Applied parts, i.e. equipment making
physical contact with the patient, comprise
nebulizer patient unit and cable and the
Ventilator Breathing System described in
System Flow Chart, Ventilation, Patient
Connection, part no. 66 92 522.
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual