11.9 ALARMS
11.9.1 ALARM LIMITS
Parameter
Airway pressure, upper limit (cmH
Airway pressure, upper limit (cmH
7
NIV
Apnea time to alarm (s)
End expiratory pressure, high limit
8
(cmH
O)
2
End expiratory pressure, lower limit
(cmH
O)
2
Expired minute volume, lower limit (l/min)
Expired minute volume, upper limit (l/min)
Respiratory rate, lower limit (b/min)
Respiratory rate, upper limit (b/min)
O
concentration high
2
O
concentration low
2
Gas supply
High continuous pressure
6. If P
rises 6 cmH
O above the set limit or if system pressure exceeds 117 ±7 cmH
aw
2
7. If P
rises 6 cmH
O above the set limit or if system pressure exceeds 117 ±7 cmH
aw
2
8. Setting the alarm limit to 0 (zero) is equivalent to turning off the alarm.
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
Factory set
default
6
O)
40
2
O) in
20
2
20
2
10
5.0
40.0
5
30
-
-
<2.0 kPa x 100 or
>6.0 kPa x 100
-
Audio Off (only
Setting range
for NIV)
16 - 120
No
16 - 40
No
15 - 45
No
0 - 47
Yes
0 - 55
Yes
0.5 - 40.0
Yes
0.5 - 60.0
Yes
1 - 160
Yes
1 - 160
Yes
Set value + 5 vol%
No
Set value - 5 vol%
No
or 18 vol%
-
-
Obstruction leading
No
to constant high
airway pressure
(>PEEP +15
cmH
O) during:
2
> 2 breaths or 5
seconds,
whichever is
greater,
15 ±1.5 s if less
2 breaths are
triggered
O, the safety valves opens.
2
O, the safety valves opens.
2
|
Technical data
| 11 |
153