4.8 START VENTILATION
The Start/Standby key is used to start and
stop both invasive and non invasive ventilation.
It is also possible in Standby to start
ventilation by pressing the Start ventilation
touchpad on the screen.
WARNING! Make sure that ventilation is
started when a patient is connected to the
ventilator. When the ventilator is in Standby,
a flashing message, Patient not ventilated,
is displayed as a reminder directly above
the word Standby.
4.8.1 START INVASIVE VENTILATION
SVS-5020_XX
1. When the system is configured for
invasive ventilation press the
Start/Standby key (1) to start ventilation,
or press the Start ventilation touchpad on
the screen.
54
|
4.8.2 START NON INVASIVE VENTILATION
(NIV) (OPTION)
00:14
1. When the Start/Standby key is pressed
and the SERVO-s Ventilator System is
configured for NIV, a waiting position
dialog is shown.
Note: All patient-related alarms are
turned off for 120 seconds.
2. Press the Start ventilation touchpad.
Note: Ventilation starts automatically
upon patient effort.
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
69;
B( 1