6.10 BI-VENT
6.10.1 FUNCTIONAL DESCRIPTION
SVS-5056_XX
Bi-Vent is pressure controlled breathing that
allows the patient the opportunity of
unrestricted spontaneous breathing. Two
pressure levels are set together with the
individually set duration of each level.
Spontaneous breathing efforts can be assisted
by pressure support
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
|
Ventilation, modes and functions
The following parameters are set:
1. Pressure high (PHigh) for the higher
pressure level (cmH
2. PEEP for the lower pressure level (cmH
3. Oxygen concentration (%)
4. Time at the higher pressure (THigh) level
(s)
5. Time at the lower pressure (TPEEP) level
(s)
6. Inspiratory rise time (s)
7. Trigg. Flow / Trigg. Pressure
8. Inspiratory Cycle-off (%)
9. PS (Pressure Support level) above PHigh
(cmH
O)
2
10. PS (Pressure Support level) above PEEP
(cmH
O)
2
SVS-5057_XX
O)
2
2
103