Alarms
|
7.2 HANDLING ALARMS
The system can signal four types of alarm:
High Priority—red background
Medium Priority—yellow background
Low Priority—yellow background
Technical—a numeric code
The following sections provide general
information about viewing, responding to,
silencing, and turning off alarms.
122
7.2.1 VIEWING THE CURRENT ALARMS
WINDOW
SVS-6000_XX
If more than one alarm is active, view the
Current Alarms Window by:
1. Press the bell(s) in the alarm message
touchpad.
All alarms (up to 10 listed by priority) are
shown in a dynamic window that will be
updated if more alarms occur while the
window is open.
2. View the current alarms.
3. Press the History touchpad.
The previous 16 alarm-dependent events
are listed chronologically, with the most
recent event at the bottom.
Note: For viewing more than the latest 10
alarms, use the Event log to view all logged
alarms.
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual