11.18 INFORMATION REQUIRED BY
ISO 80601-2-12
11.18.1 VOLUME-CONTROLLED BREATH
TYPE
Characteristics valid for patient categories and
circuit configurations intended to provide the
indicated tidal volume, at conditions specified
in ISO 80601-2-12, table 201.103.
Maximum inaccuracy of inspired tidal
volume (V
)
T
±(4 ml + 7 % of set volume) for V
±(4 ml + 7 % of set volume) for
100 ml ≤ V
≤300 ml
T
Maximum inaccuracy of PEEP
±(1 cmH
O + 5 % of set value) for
2
≥300 ml
V
T
±(1 cmH
O + 5 % of set value) for
2
100 ml ≤ V
≤300 ml
T
Maximum inaccuracy of inspired oxygen
concentration (FiO
connection port
±(3 % + 0 % of set value) for V
±(3 % + 0 % of set value) for
100 ml ≤ V
≤300 ml
T
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
≥300 ml
T
) at the patient
2
≥300 ml
T
11.18.2 PRESSURE-CONTROLLED BREATH
TYPE
Characteristics valid for patient categories and
circuit configurations intended to provide the
indicated tidal volume, at conditions specified
in ISO 80601-2-12, table 201.104.
Maximum inaccuracy of airway pressure
(P
) at the end of the inspiratory phase
AW
±(1 cmH
O + 5 % of set value) for
2
≥300 ml
V
T
±(1 cmH
O + 5 % of set value) for
2
100 ml ≤ V
≤300 ml
T
Maximum inaccuracy of PEEP
±(1 cmH
O + 5 % of set value) for
2
≥300 ml
V
T
±(1 cmH
O + 5 % of set value) for
2
100 ml ≤ V
≤300 ml
T
Maximum inaccuracy of inspired oxygen
concentration (FiO
connection port
±(3 % + 0 % of set value) for V
±(3 % + 0 % of set value) for
100 ml ≤ V
≤300 ml
T
|
Technical data
) at the patient
2
≥300 ml
T