Cardinal 201 Installation, Technical And Operation Manual page 137

Weight transmitter
Hide thumbs Also See for 201:
Table of Contents

Advertisement

pULS= (Pulse Width)
 (Trim)
 (Zero Tolerance)
8400-M117-O1 Rev D
This setting sets the width of the generated pulse (see Pulse Target
above) in milliseconds. To generate a 1 millisecond pulse, a value of '1'
should be entered, for a 10 millisecond pulse, a value of '10' should be
entered, etc.
NOTE: If the DFC Mode is set to DIO operation, this setting is ignored as
no pulse will be generated when in DIO mode.
Press the F3/  key to show the current setting. If the setting displayed
is acceptable, press the F3/  key again to save it. Otherwise, use the
F2/ and F1/ keys to enter a new setting and then press the F3/  key
to save it. Allowable values for this parameter are from 1 to 100.
This setting controls the trim value of the filling process. This is the
amount of weight below the actual target value at which the output will be
deactivated in order to accurately arrive at the target weight.
NOTE: If the DFC Mode is set to "Off" then this setting is not applicable.
Press the F3/  key to show the current setting. If the setting displayed
is acceptable, press the F3/  key again to save it. Otherwise, use the
F2/ and F1/ keys to enter a new setting and then press the F3/  key
to save it. Allowable values for this are from 0 to Fast Target.
If , the completed fill operation will immediately discharge. The
discharge will continue until the scale weight returns to zero within the
 (Zero Tolerance) value.
Press the F3/  key to show the current setting. If the setting displayed
is acceptable, press the F3/  key again to save it. Otherwise, use the
F2/ and F1/ keys to enter a new setting and then press the F3/  key
to save it.