CALIBRATION TEST EQUIPMENT REPLACEMENT
CalibratIon Test EquIpment Chart
This chart compares TM 500 product performance to that of older Tektronix equipment. Only those
characteristics where significant specification differences occur, are listed. In some cases the new instrument
may not be a total functional replacement. Additional support instrumentation may be needed or a change in
'calibration procedure may be necessary.
Comparison of Main Characteristics
OM 501 replaces 7013
PG 501 replaces 107
PG 501 - Risetime less than 3.5 ns into
107 - Risetime less than 3.0 ns into
50 O.
500.
108
PG 501 - 5 V output pulse; 3.5 ns Risetime.
108 - 10 V output pulse; 1 ns Risetime.
111
PG 501 - Risetime less than 3.5 ns; 8 ns
111 - Risetime 0.5 ns; 30 to 250 ns
Pretrigger pulse delay.
Pretrigger Pulse delay.
114
PG 501 - ±5 V output.
114 - ±10 V output. Short proof output.
115
PG 501 - Does not have Paired, Burst, Gated,
115 - Paired, Burst, Gated, and Delayed
or Delayed pulse mode; ±5 V dc
pulse mode; ±10 V output.
Offset. Has ±5 V output.
Short-proof output.
PG 502 replaces 107
108
PG 502 - 5 V output
108 - 10 V output.
111
PG 502 - Risetime less than 1 ns; 10 ns
111 - Risetime 0.5 ns; 30 to 250 ns
Pretrigger pulse delay.
Pretrigger pulse delay.
114
PG 502 - ±5 V output
114 - ±10 V output. Short proof output.
115
PG 502 - Does not have Paired, Burst, Gated,
115 - Paired, Burst, Gated, Delayed
&
Un-
Delayed
&
Undelayed pulse mode;
delayed pulse mode; ±10 V output.
Has ±5 V output.
Short-proof output.
2101
PG 502 - Does not have Paired or Delayed
2101 - Paired and Delayed pulse; 10 V
pulse. Has ±5 V output.
output.
PG 506 replaces 106
PG 506 - Positive-going trigger output signal
106 - Positive and Negative-going trigger
at least 1 V; High Amplitude out-
output signal, 50 ns and 1 V; High
put, 60 V.
Amplitude output,100 V.
067-0502-01
PG 506 - Does not have chopped feature.
0502-01 - Comparator output can be alter-
nately chopped to a reference
voltage.
SG 503 replaces 190,
190A,190B
SG 503 - Amplitude range 5 mV to 5.5 V p-p.
190B - Amplitude range 40 mV to 10 V p-p.
191
SG 503 - Frequency range 250 kHz to 250 MHz.
191 - Frequency range 350 kHz to 100 MHz.
067-0532-01
SG 503 - Frequency range 250 kHz to 250 MHz.
0532-01 - Frequency range 65 MHz to 500 MHz.
TG 501 replaces 180,
180A
TG 501 - Marker outputs,S sec to 1 ns.
180A - Marker outputs,S sec to 1 /-Is.
Sinewave available at 5, 2, and 1 ns.
Sinewave available at 20, 10,
Trigger output - slaved to marker
and 2 ns. Trigger pulses 1, 10,
output from 5 sec through 100 ns.
100 Hz; 1, 10, and 100 kHz.
One time-mark can be generated at a
Multiple time-marks can be
time.
generated simultaneously.
181
TG 501 - Marker outputs,S sec to 1 ns. Sine-
181 - Marker outputs, 1, 10, 100, 1000,
wave available at 5, 2, and 1 ns.
and 10,000 /-IS, plus 10 ns sinewave.
184
TG 501 - Marker outputs,S sec to 1 ns. Sine-
184 - Marker outputs,S sec to 2 ns. Sine-
wave available at 5, 2, and 1 ns.
wave available at 50, 20, 10, 5,
Trigger output - slaved to marker
and 2 ns. Separate trigger pulses
output from 5 sec through 100 ns.
of 1 and.1 sec; 10, 1, and.1 ms;
One time-mark can be generated at
10 and 1
/-IS.
Marker amplifier pro-
a time.
vides positive or negative time
marks of 25 V min. Marker
intervals of 1 and .1 sec; 10, 1,
and .1 ms; 10 and 1/-1s.
2901
TG 501 - Marker outputs,S sec to 1 ns. Sine-
2901 - Marker outputs,S sec to 0.1 /-Is.
wave available at 5, 2, and 1 ns.
Sinewave available to 50, 10,
Trigger output - slaved to marker
and 5 ns. Separate trigger pulses,
output from 5 sec through 100 ns.
from 5 sec to 0.1 /-Is.
One time-mark can be generated at
Multiple time-marks can be gene-
a time.
rated simultaneously.
NOTE: All TM 500 generator outputs are short-proof. All TM 500 plug-in Instruments require TM SOo-Serles Power Module.