5.6.2 Manufacturer instructions – electromagnetic susceptibility
Immunity Tests
Electrostatic discharge (ESD)
IEC 61000-4-2
Radiated RF
IEC 61000-4-3
Proximity fields from RF wireless communications equipment
IEC 61000-4-3
Fast electrical transients / burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Conducted RF
IEC 61000-4-6
Power frequency (50/60 Hz) magnetic field
IEC 61000-4-8
Voltage dips, short interruptions on power supply input lines
IEC 61000-4-11
Test frequency
(MHz)
385
450
710
745
780
810
870
930
1 720
1 845
1 970
2 450
5 240
5 500
5 785
D9U001MC0-0101_16
Table 1 - IMMUNITY to RF wireless communications equipment
Band (MHz)
380 - 390
430 - 470
GMRS 460, FRS 460
704 - 787
LTE band 13, 17
800 - 960
GSM 800/900, TETRA 800,
iDEN 820, CDMA 850,
1 700 - 1 990
GSM 1800; CDMA 1900;
GSM 1900; DECT; LTE
band 1, 3, 4, 25; UMTS
2 400 - 2 570
Bluetooth, WLAN, 802.11
b/g/n, RFID 2450,
5 100 - 5 800
WLAN 802.11 a/n
Compliance level
± 8 kV for contact discharge
± 15 kV for contact discharge
3 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
See Table 1
±2 kV for power line
repetition frequency 100 kHz
± 1 kV Line-to-line
± 2 kV Line-to-ground
3 V (0,15 MHz – 80 MHz)
6 V in ISM bands between 0,15 MHz and 80 MHz
80 % AM at 1 kHz
30 A/m
0 % UT; 0,5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270° a 315°
0 % UT; 1 cycle and 70 % UT; 25/30 cycle
Single phase: at 0°
0 % UT; 250/300 cycle
Service
TETRA 400
Pulse modulation 18 Hz
FM ± 5 kHz deviation 1
Pulse modulation 217 Hz
Pulse modulation 18 Hz
LTE band 5
Pulse modulation 217 Hz
Pulse modulation 217 Hz
LTE band 7
Pulse modulation 217 Hz
Modulation
Immunity Test
Level V/m
kHz sine
27
28
9
28
28
28
9
23
Need help?