Download Print this page

A.MENARINI ARKIVE SL Operator's Manual page 15

Electronic-mechanical device for loading slides to be archived and the retrieval of archived slides

Advertisement

The device must be handled with lifting equipment that is suitable to bear its weight.
If the device needs to be transported, it must be on its original pallet and under no
circumstances must it be stacked on top of other devices. If it is necessary to move
the device without its original pallet, it must first be adequately protected and then
lifted. The device must never be pushed or dragged, causing it to scrape along the
floor.
The device must be installed in a suitable room according to the specifications
indicated below:
• the room temperature must be between 15°C and 25°C;
• the floor must be non-slip;
• the device must be positioned perfectly level;
• the room must have load-bearing characteristics of at least 6 kN/m
• the room must have ventilation and lighting systems as prescribed by the
regulations in force in the user's country;
• the room must be clean and free of moisture;
• the room must have switches to cut off any form of electricity upstream if
necessary (emergency shut-off);
• the room must have a preferential electricity line in "absolute continuity" coming
from a UPS to which the system can be connected;
• the room must have Ethernet or Wi-Fi coverage for access to internet, which
will be used for remote control of the system, for data backup and for connecting
to the Laboratory Information System (LIS);
• the walls immediately behind the device must be fireproof and isolated from
possible heat sources.
"ARKIVE SL FINDER" OPERATOR'S MANUAL
THIS MANUAL IS THE PROPERTY OF THE MANUFACTURER AND ANY REPRODUCTION, EVEN PARTIAL, IS PROHIBITED
Approved
11/2019 - Rev. 1.3
- 15
58 -
of
2
;
.