Hainbuch TESTit Operation Manual page 48

Hide thumbs Also See for TESTit:
Table of Contents

Advertisement

TESTit
Installing and setting up the software
48
The value 1 corresponds to 100%, the value 1.05 corre-
sponds to 105%.
The Fig. presents the values 1.05 for the force and 1.02 for
the speed.
If a new centrifugal force curve is recorded under »Force-
Speed Diagram« or if an already stored centrifugal force
curve is added to the diagram, then the maximum values
for the force and speed of the diagram will be determined
with due consideration of the factor that is set.
The maximum measured force value of a centrifugal force
curve is multiplied by the factor that is set, which then
serves as the maximum value for the scaling of the dia-
gram.
Example:
A centrifugal force curve has a maximum measured
force value of 71 kN and the set factor of the force is
1.05.
The diagram shows a maximum force value of
��
max.Diagramm
= 71 ���� ∗ 1.05 = 74,55 ���� ≈ 75 ����
Specifically, the factor indicates that the diagram
shows a 5% greater value.
If several curves are now added, the curve with the maxi-
mum measured force value is always used for the scaling
of the diagram.
The same procedure is also used for the scaling of the
speed.
These settings are the basic settings.
The display can be adjusted at any time via »Select sec-
tion«.
Order hotline +49 7144.907-333
= ��
∗ ������������
max.Messkraft
Kraft