Dräger Fabius GS Supplement Manual page 15

Operator’s instruction manuals for the optional heated breathing system
Hide thumbs Also See for Fabius GS:
Table of Contents

Advertisement

Operation of HBSPS
HBSPS Problem Resolution
Note:
Call your local Authorized Service
Organization or DrägerService if the
following steps do not resolve HBSPS
problems.
Bottom of COSY Not Warm After 30 Minutes
1. Ensure that the HEATER POWER switch
(1 in
Figure
15) is in the ON position.
2. Ensure that the "HEATER POWER ON" LED is
illuminated (2 in
Figure
3. Check that the connector (1 in
cable running to the COSY is plugged in
properly.
4. Check that the MAINS POWER LED
(1 in
Figure
17) is illuminated and that the Mains
Power cord (2 in
Figure
functioning wall outlet.
"HEATER POWER ON LED" Is Not Illuminated
If the "HEATER POWER ON" LED (2 in
not illuminated when the HEATER POWER switch
(1 in
Figure
15) is in the ON position and the MAINS
POWER LED (1 in
Figure
the fuse (3 in
Figure
15). If it is blown, replace it with
the properly rated fuse.
MAINS POWER LED Is Not Illuminated
If the MAINS POWER LED (1 in
illuminated when the mains power cord
(2 in
Figure
17) is plugged into a wall outlet:
1. Check that the wall outlet is functioning and the
plug is plugged into it properly.
2. Check that the power supply-end cord plug (4 in
Figure
17) is plugged into the power supply
properly.
3. Check the two fuses in the power inlet assembly
(3 in
Figure
17). Replace them if they are blown.
Fabius GS/GS premium/Tiro/OS Supplement
15).
Figure
1) on the
17) is plugged into a
Figure
15) is
17) is illuminated, check
Figure
17) is not
Supplement #9038278, Rev. 07
Figure 15. HBSPS Front
Figure 16. HBSPS Connector
Figure 17. HBSPS Back