Alarm Parameters - Sanyo MCO-19AIC(UV) Instruction Manual

Co2 incubator
Hide thumbs Also See for MCO-19AIC(UV):
Table of Contents

Advertisement

ALARM PARAMETERS

1. Press the MENU Key from the Top Display to display the menu, select Tools, and press the ENTER
Key.
T e m p
3 7 . 0
1
2
3
4
S t a t u s
5
O K
6
7
2. Select Alarm Setting from the Select Tools Display, press the MENU Key to display the menu, select
OK, and press the ENTER Key.
S e l e c t
T e m p / C O 2
A l a r m
L C D / D A Q
U V
S e t t i n g
K e y
D a t e
3. The Alarm Setting Display will appear. The temperature alarm, CO
and door delay can be set on this display. The alarm buzzer can be silenced by pressing the BUZZER
Key. The buzzer will sound again after the specified ring back time if the condition that caused the alarm
continues. The ring back time can be set.
A l a r m
1
T e m p
3
C O 2
4
A l a r m
5
R i n g
D o o r
6
Temperature alarm: ±1.0 to ±5.0
CO
alarm: ±0.5 to ±5.0% (Default: ±1.0%)
2
Alarm delay: 0 to 15 minutes (Default: 15 minutes)
Ring back time: 1 to 99 minutes, or OFF (Default: 30 minutes)
Door alarm delay: 1 to 30 minutes (Default: 2 minutes)
25
o
C
C O 2
T o o l s
( 1 / 2 )
C a l i b r a t i o n
S e t t i n g
S e t t i n g
L o c k
P W
S e t t i n g
T i m e
S e t t i n g
A l a r m
±
1 . 0
A l a r m
±
0 . 5
D e l a y
1 5 m i n
B a c k
3 0 m i n
D e l a y
o
C (Default: ±1.0
5 . 0 %
alarm, alarm delay, ring back time,
2
o
C
( ± 1 . 0 ° C
%
( ± 0 . 5 %
( 0
-
1 5 m i n )
( 0 . O F F
2 m i n
( 1
-
3 0 m i n )
o
C)
116
M E N U
S e t
L o g
T o o l s
M E N U
O K
S v c
C a n c e l
M E N U
± 5 . 0 ° C
O K
± 5 . 0 %
C a n c e l
1
9 9 m i n )