Alarm Setting (Tools/Alarm Setting) - Sanyo MIR-554 Instruction Manual

Cooled incubator
Hide thumbs Also See for MIR-554:
Table of Contents

Advertisement

VARIOUS SETTING (MENU/Tools)
Alarm setting (Tools/Alarm, Opt., Low Humi.
Setting)
1. Select "Alarm, Opt., Low Humi. Setting" in the Select Tools screen. Press the menu button (MENU)
to show the menu window. Select "OK", and press the enter key (ENTER).
S e l e c t
L o g
D a t e
A l a r m , O p t . , L o w
K e y
D e f a u l t
D e l e t e
2. The Alarm, Option, Low Humidity Setting screen is displayed. On this screen, the temperature alarm
(Temp Alarm), delay time of alarm (Alarm Delay), alarm resume time (Ring Back) and lamp (Lamp) can
be set. The alarm buzzer is silenced by pressing the alarm buzzer stop key (BUZZER) during alarm
condition. The buzzer will be activated again after certain suspension if the alarm condition continues.
The suspension time (ring back) can be set.
A l a r m , O p t i o n , L o w
T e m p
A l a r m
R i n g
I n n e r
L a m p
L o w
H u m i d i t y
The settable range:
■ Temperature alarm (Temp Alarm): ±1.0∼±5.0
■ Delay time of alarm (Alarm Delay): 0∼15 minutes.
■ Suspension time (Ring Back): 1∼99 minutes, or OFF
3. When optional inner door (MIR-55ID) is installed, set the inner door (Inner Door) to 1.
・When the inner door is installed, the set temperature is limited under 50
・The inner door might be damaged by running at the temperature that exceeds 50
inner door (Inner Door) is 0.
4. When optional light add-on kit (MIR-L15) is installed, set the lamp (Lamp) to 1.
・When optional light add-on kit (MIR-L15) is installed, except when the set value of temperature is
o
between +2∼+50
o
+50
C after Lamp is set to 1.
5. The condensation in the chamber is reduced by running control with the set temperature between +20
o
and +40
C when the low humidity mode (Low Humidity Mode) is set to 1.
・The condensation can be found on the inside of the door or clearance in the chamber when running in
humid chamber status without low humidity mode (Low Humidity Mode is set to 0).
CAUTION
It is not possible to change the delay time of door alarm, 2 minutes only. When door is open, Open is
displayed on the top
T o o l s
M I R - 5 5 4
T i m e
H u m i .
L o c k
P W
S e t t i n g
S e t t i n g
U s e r
D a t a
H u m i d i t y
o
A l a r m
±
2 . 5
D e l a y
1 5 m i n
B a c k
3 0 m i n
D o o r
0
( 0 . N o r m a l
0
( 0 . N o r m a l
M o d e
C, the lighting is turned off. The set value of temperature is limited between +2∼
screen.
M E N U
O K
S v c
S e t t i n g
C a n c e l
S e t t i n g
C
( ± 1 . 0 ° C
-
± 5 . 0 ° C )
( 0 - 1 5 m i n )
( 0 . O F F
1 - 9 9 mi n )
1 . O p t i o n )
1 . O p t i o n )
0
( 0 . N o
1 . Y e s )
o
C.
39
o
C.
o
C as the setting of the

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents