Power and environmental ratings
For indoor use only.
100 V ac to 240 V ac, 50 Hz to 60 Hz
Power supply
(autosensing)
240
Maximum VA
Maximum 2000 m (6562 ft) above
Operating altitude
sea level
0 °C to 35 °C at up to 70% relative humidity;
Operating
temperature
at 35 °C to 50 °C, derate 3% relative humidity
per °C
–25 °C to 65 °C
Storage temperature
1 or 2
Pollution degree
Voltage: 40 V dc maximum
DC source input
electrical ratings
Current: 3 A maximum at 6 V dc, 1 A maximum
at 40 V dc
Region maximums: 10 A at 20 V
Pulser feature
output, region 4
Maximum pulse width: 1.8 ms
Maximum duty cycle: 1%
Measurement category O
Measurement
input electrical
Voltage: 40 V dc maximum HI to LO
ratings
Current: 3 A at 6 V dc; 1 A maximum at
40 V dc
Impedance: Variable
CAUTION
Carefully consider and configure the appropriate
output‑off state, source levels, and compliance levels
before connecting the instrument to a device that can
deliver energy. Failure to consider the output‑off state,
source levels, and compliance levels may result in
damage to the instrument or to the device under test.
Need help?
Do you have a question about the Keithley System SourceMeter 2601B-PULSE and is the answer not in the manual?