5. Performance Verification
Bandwidth Performance Verification
should be approximately -15.3 dB, but can vary a few tenths of a dB
either way.
Display should look like
Figure
59. If it looks noticeably different, the
c
probe tip wires may not be making contact under the pinchers.
Figure 58 Setup to measure "Vout" of probe
83