Replacing Wires on the
N5441A Solder-in Probe
C A U T I O N
The N5441A solder-in probe head's tip wires can be replaced when they
are worn out or damaged. 0.005" and 0.007" wire kits are included with
each probe head. 0.005" wire can be used for attaching to small vias.
Head
To replace wires, complete the following steps:
Begin by clamping the probe head GENTLY.
1
Figure 37 Clamping the Probe Head
Remove the old wires with tweezers while reflowing the solder from
2
the probe underside.
Apply heat quickly to avoid damaging your probe.
Figure 38 Removing the Old Wire
If necessary, add a small amount of solder to the holes and apply flux.
3