2.4.1
Sensor tags
Figure 2-1:
2.4.2
Density calibration parameters (D1, D2, K1, K2, FD, TC)
Density calibration parameters are typically on the sensor tag and the calibration
certificate.
2.5
Verify mass flow measurement
Check to see that the mass flow rate reported by the transmitter is accurate. You can use
any available method.
Connect to the transmitter with ProLink III and read the value for Mass Flow Rate in the
Process Variables panel.
Postrequisites
If the reported mass flow rate is not accurate:
•
•
2.6
Verify the zero
Verifying the zero helps you determine if the stored zero value is appropriate to your
installation, or if a field zero can improve measurement accuracy.
Configuration and Use Manual
Tag on sensors
Check the characterization parameters.
Review the troubleshooting suggestions for flow measurement issues.
Quick start
9