4. Empty the syringe entirely.
5. Pump up same type and necessary amount (max. 50 ml) of distilled water by pulling the
„knob" of the syringe outwards.
6. Place the syringe back to its seating by placing the nib to its stand first then tilt back the
syringe's neck. Make sure to stand the syringe perfectly vertical. Check the position of the nib.
On the supported B.Braun Omnifix 50 ml syringe the text should be on the back side.
7. Close the service door.

5.4.3 Replacing fuses of the main power switch

1. Switch off the main power supply and disconnect the power supply cable (see chapter 1.1.4
Base Unit).
2. The fuse compartment is located above the main power supply switch at the back side of
the Pannoramic® 1000.
3. Release lock (1) by pushing the lever to the right, so the insert (2) can be pulled out.
Sep 9, 2020 – Rev. 1
Figure 36 – Location of fuse compartment on the back panel
Figure 37 – Fuse compartment
3DHISTECH Ltd.
5.4.2 Refilling the pipetting unit
63(74)