11 INSTALLATION INSTRUCTIONS

When installing the pump make sure that accident prevention regulations and safety instructions
including those for subsequent operation are observed. The safety instructions in section must be
observed.
The following ambient conditions must be observed:
Ambient temperature during operation: max. +50 °C [122 °F]
Protect the pump against water and dust.
During operation, a sufficient supply of cooling air must be guaranteed.
Danger
Note
Danger

11.1 MOUNTING INSTRUCTIONS FOR SR25.2 PERISTALTIC PUMP

Make sure to mount the peristaltic pump SR25.2 to a front plate with a minimum thickness of 3.5 mm
[0.138"]. If the front plate is thinner than 3.5 mm [0.138"], use shims to adjust spacing. The maximum
thickness of the front plate is 5.5 mm [0.216"].
The minimum distance avoids damages to the pump motor, and the maximum distance prevents the
motor shaft from loosing the guide inside the pump head.
www.mc-techgroup.com
Do not operate the peristaltic pump SR25.X-X in hazardous areas.
The pump must only be used in the conditions specified in the technical data.
The pump should be installed away from heat sources and freely ventilated
to prevent any accumulation of heat.
For outdoor installation, the pump must be installed in a housing protected
from frost in the winter and sufficiently ventilated in summer. Exposure to
direct sunlight must be avoided.
It is therefore essential to provide protection for persons against contact with
alive parts (e.g. electrical connections, motor windings) and moving parts
(e.g. fan). Protection against the entry of foreign bodies must also be
provided.