Hart Communication With Supply Via Isolating Power Supply - Siemens SITRANS TH100 Operating Instructions Manual

Temperature transmitter with 4to 20 ma/hart
Table of Contents

Advertisement

4.3.9.2

HART communication with supply via isolating power supply

Only intrinsically safe HART communicators or HART modems are allowed to be used with an intrinsically safe power
supply.
HART communication via HART jacks of isolating power supply
Load ≥ 250 Ω only relevant if HART communication takes place via this branch. Otherwise, load of 0 to 650 Ω for ver‐
sions
or
Figure 4-17
HART communication with supply via isolating power supply
SITRANS TH100/TH320/TH420/TR320/TR420/TF320/TF420 with 4 to 20 mA/HART
Operating Instructions, 05/2020, A5E41864807-AB
Description
4.3 SITRANS TR320/TR420
51