To configure SPM, refer to the Rosemount 3051 F
Reference
Procedure
1. Select AI Block Unit Setup.
2. Place the AI Block in Out of Service mode.
3. Select the signal conditioning L_TYPE from the menu.
4. Set the XD_SCALE to the 0% and 100% scale points (the transmitter
range).
5. If the L_TYPE selected is Direct, the AI block may be placed in AUTO
mode to return the device to service.
6. If the L_TYPE selected is Indirect or Indirect Square Root, set
OUT_SCALE to change engineering units.
7. To change damping, select Change Damping.
8. Enter the desired damping value in seconds.
The permitted range of values is 0.4 to 60 seconds.
9. To configure the LCD display (if installed).
10. Select Local Display Setup.
11. Check the box next to each parameter to be displayed to a maximum
of four parameters.
Quick Start Guide
Manual.
Select Direct for pressure measurements using the device default
units.
Select Indirect for other pressure or level units.
a) Select the XD_SCALE_UNITS from the menu.
b) Enter the XD_SCALE 0% point.
This may be elevated or suppressed for level applications.
c) Enter the XD_SCALE 100% point.
This may be elevated or suppressed for level applications.
a) Select OUT_SCALE UNITS from the menu.
b) Set the OUT_SCALE low value.
This may be elevated or suppressed for level applications.
c) Set the OUT_SCALE high value.
This may be elevated or suppressed for level applications.
d) If the L_TYPE selected is Indirect, the AI Block may be placed
in AUTO mode to return the device to service.
Quick Start Guide