MULTICAL® 603
Temperature measurement
603‐A
2‐Wire Pt500
603‐B
4‐Wire Pt500
Measuring
range
603‐C/E/F
2‐Wire Pt500
603‐D/G
4‐Wire Pt500
Offset adjustment
Max cable lengths
Max ø6 mm
cable
Flow measurement V1 and
V2
CCC‐code
EN 1434 pulse class
Pulse input
Pulse ON
Pulse OFF
Pulse frequency
Integration frequency
Electrical isolation
Max cable length
Max cable length with
Cable Extender Box, Type
66‐99‐036
14
t1
t2
Inlet
Outlet
± 0.99 K joint zero point adjustment for t1, t2 and t3
See paragraph 7.3
Note: The offset adjustment is only active on measured temperatures. If, for
example, t3 has been selected for a preset value, the offset adjustment will not
influence the preset value.
Pt100, 2‐wire
2 x 0.25 mm
2 x 0.50 mm
2 x 1.00 mm
Pt100, 4‐wire
4 x 0.25 mm
Reed switches
‐ULTRAFLOW
V1: 9‐10‐11 and V2: 9‐69‐
V1: 10‐11 and V2: 69‐11
11
1xx‐2xx‐4xx‐5xx‐8xx
0xx
IC
IB
680 k pull‐up to 3.6 V
680 k pull‐up to 3.6 V 680 k pull‐up to 3.6 V
0.4 V for 1 ms
0.4 V for 300 ms
2.5 V for 4 ms
2.5 V for 100 ms
128 Hz
1 Hz
1 Hz
1 Hz
No
No
10 m
10 m
30 m
30 m
t3
t4
Control
Extra
0.00...185.00 C
(t1 and t2: Approved for 2.00...180.00 C)
2
: 2.5 m
2 x 0.25 mm
2
: 5 m
2
: 10 m
4 x 0.25 mm
2
: 100 m
FET switches
V1: 10‐11 and V2: 69‐11
9xx
IB
0.4 V for 30 ms
2.5 V for 70 ms
8 Hz
1 Hz
No
10 m
30 m
Kamstrup A/S ∙ Technical description ∙ 5512‐2029_A1_GB_06.2017
(t1‐t2)
(t2‐t1)
Heat
Cooling
measurement
measurement
Pt500, 2‐wire
2
: 10 m
Pt500, 4‐wire
2
: 100 m
24 V active pulses
V1: 10B‐11B and V2: 69B‐79B
2xx and 9xx
(IA)
12 mA at 24 V
4 V for 3 ms
12 V for 4 ms
128 Hz
1 Hz
2 kV
100 m
‐
t5
Pre‐setting
for A1 and A2