1. Unscrew both Allen screws (2) from the holes (1) located at the sides by using an Allen
screwdriver (3).
2. Carefully push the cover (5) horizontally frontward with light pressure until it can be
removed completely from the basic unit (4).
3. When you have finished maintenance or repair work, reattach the cover. Push it horizontally
towards the basic unit (4) until the two protruding guide pins (6) fit into the corresponding
recesses of the basic unit.
4. Note that the rear part of the cover must be pushed over the top part of the blue basic unit
(7).
5. To lock the cover in place, insert the screws again and fasten them.
Make sure that no cables are squeezed as you replace the cover(s).

5.3.2 Replacing the halogen lamp

March 13, 2012 - Rev. 1
Caution!
Figure 42 – Replacing the halogen lamp