7-5 Input commands (Optional)
Connect the indicator and computer. Set Parameter P09 to "PC" and P11 to "Key". Run serial
port debugging program on the computer. Input the capital number "Z", "T", "R" in the sending
area, and the indicator can conduct the corresponding actions and have key sound at the same
time.
Z=zero
T=tare
R=print

8. Error message and trouble shootings

Error
Problem
Message
ERR0
Exceed the zero range
Model
setting
ERR1
should be within 300-300000
Initial zero point exceeds 30% of full
ERR2
load
ERR3
Exceed the A/D resolution range
ERR4
EEPROM failure
Overload condition
ERR5
ERR6
Exceeds the display range
ERR8
Lower limit is higher than upper limit Reset the weight limit value.
ERR9
Exceed tare or pre-tare range
ERR10
Wrong calibration weights
www.tanhiepthanh.com.vn
Solution
The item on the pan should be within 2% of
full load.
error.
Resolution
Adjust or reset the capacity first then adjust
resolution
1.Remove the obstacle article under the pan
2.Replace the load cell or contact the
maintenance department.
1. Replace A/D
2..Replace the load cell or contact the
maintenance department.
Re-weld
maintenance department.
Remove the overload item
-----------------------------------------------
0
Place the right test weights and the
calibration value should be below full load.
JWI-3000
EEPROM
T are value≤ full load
or
contact
the
14