stable
S table print
:
CAN DIEN TU TAN HIEP THANH
key
M anual print by pressing key
:
14.
Print format
See the appendix (more than 100 formats. The appendix just shows two formats.)
15.
Filtering setting
Set the filtering level in which the stable indication turns on. The higher the setting, the slower
stabilization time
Options: 1
2
3
,
,
,
16.
Tare/Zero condition
stable
Only after the stable indication appears, Tare/Zero function acts after pressing
:
TARE
down key
always
Tare/Zero function acts by pressing down key
:
stable
auto: Press down key
after stable
17.
weight memory parameters.
ON
display the previous weight when powering on again
OFF not display the previous weight when powering on again
18.
parameter initialization
UNIT/SET
Press
6 Calibration
Note: Before calibration, please set the capacity first. The unit used in calibration is the one that
has been set before. During the calibration procedure, LONG press
normal weighing mode without saving.
Here we take 3kg/10g as an example
TARE/
1. Press and hold
2. With no load on the weighing pan, press
3. Wait till the window displays the first calibration value. (the window displays 1.000kg)
Note: The first calibration value is default. If the capacity has been changed, the default
value is 1/3 of full load.
www.tanhiepthanh.com.vn
w eight is equal or more than 20d
(
PRINT
4
ZERO
or
TARE
ZERO
or
twice to begin initialization until the window displays
while powering on. Do not release it till the window displays
)
TARE
even if it is not stable, but Tare/Zero function acts
TARE/
to start zero point calibration.
JWI-3000
ZERO
or
even if it is not
OK"
"
ZERO/ESC
to return to
CAL"
"
10