Fluke 80i-110s User Manual page 21

Ac/dc current probe
Hide thumbs Also See for 80i-110s:
Table of Contents

Advertisement

80i-110s Instruction Manual
AC Measurement:
5100B
SETTINGS
0.1A, 60 Hz
0.3A, 400 Hz
0.5A, 2 kHz
0.6A, 4 kHz
Table 3. Performance Test Points (5100B and 5220A)
Currrent Range 0 to 100A
DC Measurement:
5100B
SETTINGS
1V
3V
5V
7V
9V
AC Measurement:
5100B
SETTINGS
1V, 5 kHz
2V, 5 kHz
3V, 2 kHz
5V, 400 Hz
7V, 60 Hz
RMS AMPS
MEASURED
1A
3A
5A
6A
DC AMPS
MEASURED
10A
30A
50A
70A
90A
RMS AMPS
MEASURED
10A
20A
30A
50A
70A
LOW LIMIT
OUTPUT
92 mV
286 mV
465 mV
559 mV
LOW LIMIT
OUTPUT
95.5 mV
287.5 mV
439.5 mV
615.5 mV
765.0 mV
LOW LIMIT
OUTPUT
92.5 mV
185.5 mV
254.5 mV
439.5 mV
595.0 mV
19
HIGH LIMIT
OUTPUT
108 mV
314 mV
535 mV
641 mV
HIGH LIMIT
OUTPUT
104.5 mV
312.5 mV
560.5 mV
784.5 mV
1035.0 mV
HIGH LIMIT
OUTPUT
107.5 mV
214.5 mV
345.5 mV
560.5 mV
805.0 mV

Hide quick links: