18
Table 2. Performance Test Points (5100B Direct)
Current range 0 to 10A
DC Measurement:
5100B
SETTINGS
0.1A
0.5A
0.9A
Figure 6.
Toroid Coil Construction
DC AMPS
MEASURED
1A
5A
9A
80i-110s Instruction Manual
LOW LIMIT
OUTPUT
92 mV
480 mV
868 mV
HIGH LIMIT
OUTPUT
108 mV
520 mV
932 mV