3. After two seconds the unit will enter normal operation and display the current temperature
of the block.
4. Press the Up or Down switches to set the block to the desired temperature. The display
will change to the current temperature set point. The temperature is set in 0.1 degree C
increments. Holding the Up or Down key for more the eight seconds will increase or
decrease the set point rapidly. The display will automatically return to current temperature
four seconds after the switch is released. Note, the temperature may not be adjusted with
the unit actively heating the block.
5. The temperature setting will be stored in memory when the display returns to the current
block temperature. This setting will be retained if the power is turned off. Note the display
must return to current temperature before the power is turned off or the new set
temperature will NOT be saved.
6. Press the START/STOP to begin the temperature control. The "^" will be displayed on the
LCD when the unit is in active temperature control.
7. The LED will turn on and off with the heating element.
8. The LED will be off when the unit is turned off and when the heater is not actively heating
the block.
9. The maximum set temperature is 101C.
10. The Mode switch will scroll through current temperature, the elapsed time from the press
of the START button, LCD off and countdown timer.
Heating & Countdown Timer
1. Turn the unit on using the Main Power rocker switch.
2. Press the Up or Down switches to set the block to the desired temperature. Holding the
Up or Down key for more the eight seconds will increase or decrease the set point
rapidly. The display will automatically return to current temperature four seconds after the
switch is released. Note, the temperature may not be adjusted with the unit running.
3. Press the Mode button until "1:00" is displayed. This is the default Count Down timer
setting. The maximum time is setting is 19:59.
4. Press the up or down key to set the desired count down time.
5. Press Start/Stop button to begin heating.
6. The LCD will alternate between actual temperature and time remaining.
7. When time runs out the LCD will alternate between the current temperature and "---". The
beeper will sound when "---" is displayed.
8. Heating will NOT turn off when time expires.
9. To cannel the countdown timer press the Mode key one time to advance to the standard
temperature only read out.

6.0 Calibration

1. Press and hold the Start/Stop then simultaneously powering up the unit with the
ON/OFF switch.
2. Release the Start/Stop button when the LCD displays "---".
3. The will beep 4 times and the display will show the calibration set point and the ":" will
flash rapidly.
4. Press the UP or Down keys to set the desired temperature at which you want to
calibrate the unit.
5. Press the Start/Stop to begin calibration. The "^" will be displayed on the LCD when
the unit is in active temperature control.