SAFETY PRECAUTIONS

Supplement for medical Use

MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and needs to be installed and put into service
according to the EMC information provided in the Accompanying documents.
Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL EQUIPMENT.
WARNING : The use of ACCESSORIES, transducers and cables other than those specified, with the exception of transducers
and cables sold by the manufacturer of the EQUIPMENT or SYSTEM as replacement parts for internal components,
may result in increase EMISSIONS or decreased IMMUNITY of the EQUIPMENT or SYSTEM.
Information of Cables and Accessories
Make sure to use specified Accessories as described below, in order to complied with IEC60601-1-2.
Note : Please refer to a part of "SAFETY PRECAUTIONS" (page 4) and "Accessories" (page 90) in the Instructions.
WARNING : The EQUIPMENT or SYSTEM should not be used adjacent to or stacked with other equipment and that if adjacent
or stacked use is necessary, the EQUIPMENT or SYSTEM should be observed to verify normal operation in the con-
figuration in which it will be used.
WARNING : The use of ACCESSORY, transducer and cable with EQUIPMENT and SYSTEM other than those specified may
result in increase EMISSIONS or decreased IMMUNITY of the EQUIPMENT or SYSTEM.
Guidance or Manufacturer's declaration - electromagnetic emissions
The Model BD-X201M is intended for use in the electromagnetic environment specified below.
The customer or the user of the Model BD-X201M should assure that it is used in such an environment.
Emission Test
RF emissions
CISPR11/EN55011
RF emissions
CISPR11/EN55011
Harmonic emissions
IEC/EN 61000-3-2
Voltage Fluctuations/
Flicker emissions
IEC/EN 61000-3-3
6
Compliance
The Model BD-X201M uses RF energy only for its internal func-
Group 1
tion. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
The Model BD-X201M is suitable for use in all establishments,
Class B
including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
Class A
buildings used for domestic purposes.
Complies
Electromagnetic environment - guidance