Table 2-7
Operation of the M1971 Series mixer's path AUX Equipment
Step
Action
a. Choose the Normal path.
b. Apply a signal to the mixer input.
Tune the analyzer
c. N9030A/B, N9020A/B, and N9010A/B:
N9030A/B, N9020A/B, and N9010A/B:
Press Input/Output, External Mixer, Mixer Path, and
Turn on the Aux
choose Aux Equipment.
Equipment path
N9040B: Press Input/Output, tap Mixer Path, and
choose Aux Equipment.
N9030A/B, N9020A/B, and N9010A/B:
Press Input/Output, External Mixer, Mixer Path, Aux
Select the User IF
Equipment, and set User IF Frequency.
Frequency
N9040B: Tap User IF Frequency and set the
frequency value.
42
Press FREQ Channel and enter a center frequency
or a start and stop frequency.
N9040B: Press FREQ and enter a center frequency
or a start and stop frequency.
Notes
Press Input/Output, External Mixer, Mixer Path,
and choose Normal. This is because starting in
the Normal path allows you to see the signal on
the analyzer screen which makes the initial setup
easier.
Be sure the signal is centered on the screen
since the analyzer will be set to Zero Span when
the Aux Equipment path is selected.
The signal will disappear from the analyzer
screen, since the mixer is only providing a signal
to the Aux Out port. The analyzer will be set to
Zero Span.
Default setting is 1.2 GHz.