Using The M1971 Series Mixer With The N9040B Signal Analyzer; Equipment Setup (Mixer's Path Normal Mode Or Dual Conversion Mode); Table 2-4 Equipment Setup For The M1971 Series Mixer And The N9040B Signal Analyzer - Keysight Technologies X Series User Manual

Hide thumbs Also See for X Series:
Table of Contents

Advertisement

2
Operation

Using the M1971 Series Mixer with the N9040B Signal Analyzer

Refer to
configuration requirements.
The following examples explain how to connect the external mixer to the signal
analyzer, and how to use the signal-identification functions.
For additional information regarding use with a particular X-Series analyzer, refer
to that analyzer's User's and Programmer's Reference Guide.
Equipment setup (mixer's path Normal mode or Dual Conversion
mode)
Table 2-4
Equipment setup for the M1971 Series mixer and the N9040B signal
analyzer
[a]
Action
Step
a. Connect an SMA cable
1. SMA connection
a. Connect a USB cable from
2. USB connection
a. Connect a BNC cable from
3. BNC connection
(Dual conversion
path only)
34
"Spectrum Analyzer Retrofit Requirements"
from the mixer to the
spectrum analyzer EXT
MIXER front panel SMA.
the mixer to the spectrum
analyzer.
the mixer's 10 MHz In to
the signal analyzer's rear
panel 10 MHz Out.
Notes
The torque for the SMA cables and adapters should not exceed 8 in-lbs.
The LO/IF connection between the mixer and the signal analyzer must be
in place before the USB cable from the mixer is connected. Connecting
the USB cable automatically triggers the LO adjustment, and if the LO/IF
cable is not connected the adjustment will not complete and an error will
occur.
When a connection is made, the green LED on the mixer lights up
indicating that the mixer has power and the processor inside the mixer is
running. The spectrum analyzer is automatically switched to external
mixing mode, and the mixer model number and mixer option are used to
set the start and stop frequencies for the mixing band.
The 10 MHz signal from the signal analyzer provides the LO signal to the
mixer's second converter in the dual conversion path.
on page 20 for the
Keysight M1971 Series User's Guide