Using The M1971 Series Mixer With The N9030A/B, N9020A/B, And N9010A/B Signal Analyzers (Option Exm); Equipment Setup (Mixer's Path Normal Mode Or Dual Conversion Mode); Table 2-1 Equipment Setup For The M1971 Series Mixer And The N9030A/B, N9020A/B, And N9010A/B Signal Analyzers (Option Exm) - Keysight Technologies X Series User Manual

Hide thumbs Also See for X Series:
Table of Contents

Advertisement

2
Operation
Using the M1971 Series Mixer with the N9030A/B, N9020A/B, and
N9010A/B Signal Analyzers (Option EXM)
Refer to
configuration requirements.
The following examples explain how to connect the external mixer to the signal
analyzer, and how to use the signal-identification functions.
For additional information regarding use with a particular X-Series analyzer, refer
to that analyzer's User's and Programmer's Reference Guide.
Equipment setup (mixer's path Normal mode or Dual Conversion
mode)
Table 2-1
Equipment setup for the M1971 Series mixer and the N9030A/B, N9020A/B, and
N9010A/B signal analyzers (Option EXM)
[a]
Action
Step
a. Connect an SMA cable from
the mixer to the spectrum
1. SMA connection
analyzer EXT MIXER front
panel SMA.
a. Connect a USB cable from
2. USB connection
the mixer to the spectrum
analyzer.
a. Connect a BNC cable from
3. BNC connection
the mixer's 10 MHz In to the
(Dual conversion
signal analyzer's rear panel
path only)
10 MHz Out.
26
"Spectrum Analyzer Retrofit Requirements"
Notes
The torque for the SMA cables and adapters should not exceed 8 in-lbs.
The LO/IF connection between the mixer and the signal analyzer must be
in place before the USB cable from the mixer is connected. Connecting
the USB cable automatically triggers the LO adjustment, and if the LO/IF
cable is not connected the adjustment will not complete and an error
will occur.
When a connection is made, the green LED on the mixer lights up
indicating that the mixer has power and the processor inside the mixer is
running. The spectrum analyzer is automatically switched to external
mixing mode, and the mixer model number and mixer option are used to
set the start and stop frequencies for the mixing band.
The 10 MHz signal from the signal analyzer provides the LO signal to the
mixer's second converter in the dual conversion path.