5.11
Parallel
Operation
Parallel
operation
is
possible
on
adjacent load
units
of the
same
type
in
the
frame
(up
to 5 units
on
the
PLZ-50F).
The
channel
at
the
left
end
facing
the front
of
the
connected channels
is
the master.
Other channels
are
slaves.
Channels
operating
in parallel
are
considered
a single channel.
The
total
current
and
resistance
(conductance)
are
used
for
displaying
and
setting.
Within
the
same
frame,
a
combination
of
parallel
and
single-unit
operation
is
possi-
ble.
M:
Master
S;
Slave
When
parallel
operation of
3
units
and
parallel
operation of 2
units are
combined
using
the
PLZ-50F
u
u
o
I
M:
Master
S:
Slave
U: Single load
unit
When
parallel
operation of
3
units
and
single-unit
operation of
2
units are
combined
using
the
PLZ-50F
Fig.
5-20
Combination
of
parallel
operation
and
single-unit
operation
Table 5-6
The number
of units
in
parallel
and
the capacity
Number
of Units
in
Parallel
PLZ70UA
PLZ150U
2
30
A
/
150
60A/300W
3
45
A
/
225
W
90A/450W
4
60
A/
300
W
120A/600W
5
75
A
/
375
W
150
A
/
750
W