Command
*IDN?
CAL:DEV:DATE
<year>,<month>,<day>
CAL:DEV:SI
ON|OFF
CAL:USER:ADJ<n> <value>
CAL:USER:LOW <temp>
CAL:USER:HIGH <temp>
CAL:USER:TEMP<n> <temp>
Table 4. List of All Commands
Returns the instrument identification string that
indicates the manufacturer, model number, serial
number, and code firmware version.
Sets the last calibration date for the device. The
<year> parameter is a four-digit number, 2000 to
2099. The <month> parameter is a one or two-
digit number, 1 to 12. The <day> parameter is a
one or two-digit number, 1 to 31. This command
is password protected.
Sets the SI units lock. The parameter is either
OFF or ON. This command is password
protected.
Sets the adjustment at one of the calibration
temperature points for a manual calibration.
Number <n> specifies the low-temperature
calibration point (1), mid-range calibration point
(2), or high-temperature calibration point (3). The
parameter <value> is the temperature adjustment
in degrees Celsius. This command is password
protected.
Automatically calibrates the low range of the
Product. The probe temperature must be held at
a constant, known temperature near the low end
of the range. The <temp> value is the
temperature in degrees Celsius or Fahrenheit
that the Product should be displaying after it is
calibrated. This command is password protected.
The Product temperature units must match the
units of the <temp> value.
Automatically calibrates the high range of the
product. The probe temperature must be held at
a constant, known temperature in degrees
Celsius or Fahrenheit near the high end of the
range. The <temp> value is the temperature that
the Product should be displaying after it is
calibrated. This command is password protected.
The Product temperature units must match the
units of the <temp> value.
Sets one of the calibration temperature points for
a manual calibration. Number <n> specifies the
low-temperature calibration point (1), mid-range
calibration point (2), or high-temperature
calibration point (3). The parameter <temp> is the
temperature of the calibration point in degrees
Celsius. This command is password protected.
Command Explanation
Stik Thermometer
Remote Communications