Changing Fuses
8
CAUTION
!
Before removing fuse, make sure that the main power switch is in the off position and
・
the power cord is disconnected from the outlet.
Make sure and use only the proper type and rated fuse.
・
Failure to use the proper type and rated fuse will result in fire, injury, electric shock
※
and/or product damage.
・Push on the clips on the left and right of the fuse holder, and remove the fuse holder and fuses.
・Remove the bad fuse or fuses and replace with the proper type and rated fuse.
・There is the following fuse depending on the input power source voltage.
T3.15AL(250V) Order No. : U195-152
・Replace the fuse holder containing the fuses into the fuse inlet box and make sure it snaps into place.
Power Cord Connection
9
Insert the female plug into the main power inlet box
CAUTION
!
Two types of power source cords are available: For 115V &230V. Select the proper cord
・
according to the power source being used. The type of power source cord for 230V
should be specified when the system is ordered.
Only use grounded power sources. Failure to properly ground the unit may result in
・
electric shock, injury, fire and/or damage to the system components.
9
Pull out, pushing on the clips
left and right
Replace and get back
Fig.5
⑦
in the right side of the unit.
Fig.6