Operation Procedures
12
MANUAL Control mode (Fig.9, 10)
●
(1)Turn the power switch on ⑧.( The green power source lamp B lights)
(2)Push the control switch ⑪ to light MANUAL lamp F .
(3)Push the Motor Speed Selection Switch ⑫ (INT・EXT・500) to light Internal Lamp (INT).
(4)Push the FWD/REV.Switch⑩ to light FWD./REV. Lamp (Forward Rotation Lamp D or Reverse
Rotation Lamp E .)
(5)Turn the Speed Adjustment Knob fully to the left.
(6)When utilizing centering mode, push Motor Speed Selection Switch ⑫ to light 500 lamp H . And
push start switch ⑨ to rotate at 500min
7
(
)Push the START switch ⑨ and check that the Motor Lamp C lights. Adjust the Speed
Adjustment Knob by turning clockwise to select the desired motor speed. The display on the
Digital Speed Indicator ⑭ will display actual motor speed. The Digital Speed Indicator displays
speed in 2-digits: 1-80 (80 equals the max. motor speed 80,000)
(The motor speed display 1 means 1,000 min
8
(
)Pushing the Start Switch a second time will extinguish the Motor Lamp and the motor will stop.
(Reset the Speed Adjustment Knob ⑬ to the minimum speed setting, and restart the E800Z
according to the procedure mentioned in step ⑦ of this section
CAUTION
!
If an excessive load is applied to the motor and spindle, the Error Lamp A will light and
the protection system will stop the motor to protect the motor. After reducing the load,
the protect circuit can be reset. Push Start Switch
and protection circuit will be released
AUTO Control mode (Fig.9,10)
●
(1)Attach cable connectors to Input/Output connector CN1・CN2 (①・②).
(Cable connector type:Dsub25, Dsub15 and M2.6 Attachment Screw are not supplied with the
system and need to be purchased separately.)
(2)Turn the Power Switch on ⑧ (The green power source lamp B will light.)
(3)Push Control Switch⑪ to light AUTO Lamp G . Push the Motor Speed Selection Switch ⑫ to light
the EXT Control Lamp I to set external control mode.
E xternal Input/Output Connector Specifications,
Refer to ⁄3
In Auto control mode motor speed can be controlled either by the Speed Adjustment knob or an
external signal. If you desire the Speed Adjustment knob control use the ⑫ Motor Speed
Selection Switch to select INT.
If you desire external signal control select EXT.
(4)The motor rotation direction cannot be changed by Rotation Direction Input Signal (DIR̲IN), while
the motor is rotating.
11
-1
.
-1
.)
the Error Lamp A will extinguish
⑨
for proper connections.