Protection System
14  
1
Operating Condition Monitoring
Constant  monitoring  of  the  control  unit,  motor,  spindle  and  the  pressure  of  the  cooling  air  to 
detect undesired operating conditions.  This system helps prevent errors that will result in unsafe 
operating conditions. 
・Motor is stopped. 
・ERROR Lamp (ERROR)  A  will flash. 
・The error Codes (listed in Table1) will be displayed on the Digital Speed Indicator ⑭. 
    ・Trip Signal is output to External Input/Output Connector CN2: Pin8(ERR). 
2
ERROR Reset
There are 2 methods for resetting the control unit after an ERROR.
・Manual mode, push START Switch ⑨ on the front panel to reset. 
・Auto  mode,  switch  off  (Open)  Pin14  (START)  of  External  Input/Output  Connector  CN2  to 
reset →ON (Closed) Start to rotate.
Display Code 
Detector Function 
E1 
Excessive Current 
E3 
Motor Sensor 
E4 
Thermal Protection 
E7 
Low Air Pressure 
E8 
Excessive Load 
EA 
External Start UP Error 
EE 
Emergency Stop Error 
EH
Over Speed
3
Over Load Characteristics
6  LED  load  monitor  allows  the  operator  to  assess  load  conditions  easily.  Continuous  Duty  (3 
Green LEDs), Intermittent Duty (2 Yellow LEDs), Overload (1 Red LED). If the RED LED is lit for 
more  than  15  seconds,  the  overload  protection  system  will  activate  and  interrupt  power  to  the 
motor. 
When  using  E800Z  with  no  load,  LOAD  Monitor  Lamp  (Green  LED)  may  flash.  This  is  not  an 
error. (EM801, EM805, NR3080, NRA5080) 
・Error Code E8 is displayed on Motor Speed Indicator (SPEED)  ⑭. 
・Error Lamp (ERROR)  A  will flash. 
・Error Signal is output to Pin8 (ERR) of External Input/Output Connector CN2.
!  CAUTION
If the system is operated in overload repeatedly and for a long period of time, this will
cause damage to E800Z, due to overheating of the control unit, motor and spindle.
Recommended usage under load is with 3 or less Green LEDs lit: Continuous
Operation Use Area) as much as possible.
Table 2 
Excessive Current to Motor. 
Motor Connector is disconnected. 
Over 90℃  
Air pressure for motor cooling is too low. 
Continuous excessive load. 
AUTO control mode, Motor Speed Control Voltage has been supplied before Power Source is ON. 
Input Emergency Stop Signal. 
  Excessive Motor Speed(85,000min
Error 
-1
)
18