MG14 operating manual Rev 1.4
An 8-pin J-45 female appliance socket is available for each sensor. See Fig. 3-4, item D
Pin assignment
1
+24VDC supply
2
Supply GND
3
Analog Input
4
n.c.
5
Analog GND
6
ITR emission OK
7
ITR Degas
8
ITR emission ON/OFF
n.c. - not connected
!
Improper transmitter.
Transmitters which are not designed for use with the MG14 may damage the device.
Operate the MG14 with proper transmitters only. See Chapter 2.3.1 Sensor
Connections.
!
Multiple connection.
Only one transmitter may be connected to each of the channels. Otherwise the
connected transmitters will be damaged.
Never connect more than one transmitter per channel.

3.3.6 RELAYS

Fig. 3-7 Channel 2 appliance socket (RJ45 )
CAUTION
CAUTION