Before Testing; Control Solution Testing; Performing A Control Solution Test - Fora COMFORT check G40 Owner's Manual

Table of Contents

Advertisement

Available languages

Available languages

  • ENGLISH, page 1
If you want to turn off an alarm, find the alarm number by
pressing SET in the setting mode, press
to "OFF".
At the time of your alarm, the meter will beep and automatically
turn on. You can press
M
to silence the alarm and insert a test
strip to begin testing. If you do not press
for 2 minutes then switch off. If you do not want to test at this
time, press
to switch off the meter.
M
Press SET to turn off the meter. The meter repeats all your
settings before shutdown.
Congratulations! You have completed all settings!
Note:
• These parameters can ONLY be changed in the setting mode.
• If the meter is idle for 3 minutes during the setting mode, it will
switch off automatically.
M
to change from "ON'
, the meter will beep
M
EN-5

Before Testing

Control Solution Testing

FORA Control Solution contains a known amount of glucose that
reacts with test strips and is used to ensure your meter and test
strips are working together correctly.
Do a control solution test when:
you suspect the device or test strips are not working properly.
9
9
your blood glucose test results are not consistent with how
9
9
you feel, or if you think the results are not accurate,or
you have dropped or think you may have damaged the meter.
9
9

Performing a Control Solution Test

Step 1: Insert the test strip to turn on the meter.
Insert the test strip into the meter. Wait for the meter to display
the test strip and blood drop symbols.