9.0  RS‐232
INTERFACE 
 
The  GBC/GFC  Series  are  supplied  with  a  RS‐232  bi‐directional  interface.  The  scale 
when connected to a printer or computer through the RS‐232 interface, outputs the 
weight, unit weight and count. 
 Specifications: 
RS‐232 output of weighing data
ASCII code 
Adjustable Baud rate, 600, 1200, 2400, 4800, 9600 and 19200 baud 
8 data bits 
No Parity 
Connector:
9 pin D‐subminiature socket 
Pin 3  Output 
Pin 2  Input 
Pin 5  Signal Ground 
The scale can be set to print text in English, French, German or Spanish.  See the RS‐
232 parameters section for details. 
All lines end with line feed <lf>,  and carriage return <cr>characters. 