APPENDIX F.
FIRMWARE UPDATE
WARNING! Performing a firmware update on a Node might reset the Node
memory. Retrieve all data from the Network prior to performing a
firmware update.
F.1 PROCEDURE
1.
Put the Network in deployment mode (see Section 2.2).
2.
Power off the Network Supervisor by moving the battery select switch to
"OFF" (see the figure below), or by disconnecting the external battery.
FIGURE 26: Battery Selector Switch in OFF Position
3.
Ensure that the unit is completely discharged of electrical potential by
pressing the status button repeatedly until no LED lights flash.
4.
For RS-232 Supervisors:
a.
Connect the COM-108 (RS-232 cable) to the 10-pin connector on the
Supervisor.
b.
Connect the 8001-7 (USB to RS-232 adapter) to the COM-108.
c.
Connect the 8001-7 to the PC. The figure below shows the completed
connection.
8001-7 USB to
RS-232 Adapter
COM-108
RS-232 Cable
FIGURE 27: Cable Connection
For all other Nodes/Supervisors:
MODEL 8800 SERIES GEONET WIRELESS DATA ACQUISITION NETWORK | FIRMWARE UPDATE | 31
26:
27: