Labnet Prism R User Manual page 15

Refrigerated microcentrifuge
Table of Contents

Advertisement

temperature can also be changed during the run.
Pre-Cooling starts a temperature-control run at 6,000RPM.
It allows the rotor to be cooled down to the new, pre-set
temperature
temperature to 4 ° C takes approximately 8 min. (max. 16
minutes.) To activate Pre-Cooling function, press and hold
button, then press
"RPM" and "TEMP" icons will flash during Pre-Cooling. The
Pre-Cooling
(START/STOP) key. The Pre-Cooling will function only
when the set temperature is lower than the ambient
temperature.
When the centrifuge is powered on, the refrigeration
system will cool down the chamber to the pre-set
temperature before or after the run if the chamber
temperature is above the pre-set temperature.
Note:
1. To extend the life cycle of the compressor, always remember
to inactivate the Standby Cooling (by increasing the set
temperature above room temperature) or power off the
centrifuge when it is not in use.
2. When the lid is open, the Standby Cooling will stop after 3
minutes.
12
Pre-Cooling
quickly.
can
Standby Cool
Cooling
(START/STOP) key. Both
be
stopped
ng
i
down
from
by
pressing
room
the