Mindray DC-8 EXP Operator's Manual page 125

Diagnostic ultrasound system
Hide thumbs Also See for DC-8 EXP:
Table of Contents

Advertisement

Tools
Descriptions
MR Flow
Mitral Regurgitation
Rate
Flow Rate
Mitral Valve
MR Fraction
Regurgitation
Fraction
Mitral Valve
Effective
MR EROA
Regurgitant Orifice
Area
Operating Procedures
1. Enter color mode, adjust the color map until the aliasing appears.
2. Select [PISA MR] in the measurement menu.
3. Measure MR Rad using PISA caliper.
Input MR Als.Vel.
4. Measure the MR spectrum by D trace to obtain:
MR Vmax
MR VTI
MR Flow, MR Flow Rate and MR EROAare calculated automatically.
If MV SV is measured, MR Fraction will be calculated automatically.
PISA AR
Aortic valve regurgitation (PISA AR) needs to be measured in Color and Doppler mode.
Study Items
Tools
AR Rad
Aortic Valve Stenosis Radius
Aortic Valve Regurgitation
AR VTI
Velocity-Time Integral
Aortic Valve Regurgitation
AR Als.Vel
Aliasing Maximum Velocity
Study Results
Tools
Descriptions
Aortic Regurgitation
AR Vmax
Maximum Velocity
Aortic Regurgitation
AR Flow
Flow
AR Flow
Aortic Regurgitation
Rate
Flow Rate
Aortic Valve
AR Fraction
Regurgitation
Fraction
MR
Flow
Rate(ml/s)
MR
Fraction
MR
EROA(cm)
Descriptions
Obtained from AR VTI measurement
AR
Flow(ml)
AR
Flow
Rate(ml/s)
AR
Fraction
Formulae
π
=
2
MR
Rad(cm)
MR
Flow(ml)
=
(Nounit)
MV
SV(ml)
π
×
2
2
MR
Rad(cm)
=
2
MR
Vmax(cm/s)
Operations
PISA measurement
D Trace in Doppler General Measurements
You can select to use top aliasing velocity or
bottom aliasing velocity or input the value
directly.
Formulae
π
×
2
2
AR
Rad(cm)
AR
=
ARV
max(cm/s)
π
=
2
AR
Rad(cm)
AR
Flow(ml)
=
(Nounit)
AV
SV(ml)
×
2
MR
Als.Vel(cm
/s)
×
100
%
MR
Als.Vel(cm
/s)
Als.Vel(cm
/s)
×
AR
VTI(cm)
×
2
AR
Als.Vel(cm
/s)
×
100
%
Cardiology 6-33

Hide quick links:

Advertisement

Table of Contents
loading
Need help?