S
MART
The default setting for S
Turn off if the patient is experiencing false S
conduct a mask-fitting session that would create high leak between adjustments.
L
EAK ALERT
The default setting for Leak Alert is ON. The Leak Alert alarm is triggered when leak
rises above the threshold level of 30 L/min for more than 20 seconds.
Turn off if you want to conduct a mask-fitting session, or if the patient is experiencing
high leak yet wants to continue their treatment.
R
ANGE OF PARAMETERS
The following table sets out the ranges of the parameters that can be adjusted in the

Settings menu.

Parameter
ASV-CS mode
EEP (cmH
MIN PS (cmH
MAX PS (cmH
CPAP mode
CPAP (cmH
Start CPAP (cmH
Ramp (minutes)
Max ramp (minutes)
* Advanced menu
A
UTO
MAX PS minus MIN PS is greater than or equal to 5 at all times.
EEP + MAX PS will not exceed 20 cmH
To satisfy these limits:
1. Setting EEP automatically adjusts MIN PS then MAX PS.
2. Setting MIN PS automatically adjusts MAX PS.
10
S
TOP
MART
Minimum
O)
4
2
O)*
3
2
O)*
8
2
O)
4
2
O)
4
2
0
0
S
CS2 will automatically limit to ensure that:
ET
S
is ON.
TOP
S
MART
Maximum
10
6
16
13
13
45
45
O.
2
events, or if you want to
TOP
Default
5
3
10
10
4
20
30

Hide quick links: